Finishing Calibration
To finish calibrating your instrument:
- Remove all shorting blocks and connections from the instrument.
- Reset the calibration message.
- Secure the instrument with the calibration password.
- Record the new calibration count.
- Perform the performance verification tests to verify the calibration.